6 K-type non-grounded thermocouples
-100 to 1000°C
-148 to 1832°F
±0.5° C
±0.9° F
0.1° C
0.1° F
40,000 Datasets. Each Dataset consists of up to 6 channels.
2 seconds (Fixed)
Maximum recording time is 22 hours
2 AAA batteries
30 hours (recording time)
30.5 x 25.5 x 15 cm
(12" x 10" x 6")
9 kg (20 lbs.)
LOGGER KITS COME with temperature logger, stainless steel high temperature barrier box, choice of holder, 6 engraved probe ID tags, two AAA batteries, instructions, Long Form Certificate of Calibration traceable to NIST, carrying case, USB cable, PosiSoft Desktop software, and two year warranty.
LOGGER ONLY COMES with temperature logger, two AAA batteries, instructions, Long Form Certificate of Calibration traceable to NIST, USB cable, PosiSoft Desktop software, and two year warranty.
The PosiTest OTL is available in two kit configurations or as a standalone logger. The PosiTest OTL XT features an enhanced holder with heatsink for greater endurance at high temperatures.
The chart and table above represent the combined thermal characteristics of the PosiTest OTL, Barrier Box and choice of Holder/Holder XT.
Add supplier cure schedule specifications, batch annotations, label individual channels, set threshold temperatures, and compare datasets to reference batches.
Includes a customizable, templated PDF Report Generator. No internet connection required.
No. The PosiTest OTL is designed for simple operation and features an easy-to-use interface.In lieu of demonstration models, we provide unlimited technical support via telephone and/or email.
The PosiTest OTL arrives fully calibrated and ready to measure. A Long-Form Certificate of Calibration traceable to NIST is included, which documents actual readings taken by your instrument at our calibration laboratory on standards traceable to a national metrology institute. Beware of ‘Certificates’ or ‘Certificates of Conformance’ offered by competitors. These typically do not include actual instrument readings, and are often insufficient to meet common quality requirements.
The PosiTest OTL and K-type thermocouple Temperature Probes are calibrated at the factory — with no expiry or end date. They have no shelf life; therefore the length of time in use is not a good indicator for recalibration intervals.
Many organizations with quality programs and recertification programs require annual recalibrations, and this is a good starting point for most users. DeFelsko recommends that customers establish calibration intervals based upon their own experience and work environment. Based on our product knowledge, data, and customer feedback, a one year calibration interval from the date the instrument was placed in service is a good starting point.
No. Due to the lengthy calibration process, it is more economical for the PosiTest OTL temperature probes to be replaced if required.
The PosiTest OTL is available in two configurations. The PosiTest OTL XT Kit features an enhanced holder with heatsink for greater endurance at high temperatures.
A wide variety of spring clamp and magnetic temperature probes are available with either 1.8, 3.6, or 6 m (6’, 12’, or 20') stainless steel braided cables for testing air or surface temperatures. Each temperature probe includes a Certificate of Calibration traceable to NIST.
Before returning your gage or instrument for service, review the troubleshooting steps on the Troubleshooting guide, or speak with a member of our technical support team by calling +1-315-393-4450.
A wide variety of temperature probes are available for the PosiTest OTL with 1.8, 3.6, or 6 m (6', 12', or 20') stainless steel braided cables (sold separately).
Six (6) replacement engraved ID tags for PosiTest OTL temperature probes